Note: the BabyCO meter should only be connected to a computer
that is manufactured in accordance with EN60950 1992/1993 –
‘Safety of Information Technology Equipment including Electrical
Business Equipment’.
Calibration
Calibration will remain stable to within 2% over one month and
typically to within 10% over 6 months. CareFusion supplies
calibration gas (20ppm CO in air) and recommends that the unit be
recalibrated on a 6 monthly basis. See page 15 for calibration
accessories.
To carry out the calibration locate the calibration switch on the right-
hand side of the instrument as shown below.
Push the slide switch to the CO - PPM position and wait for the unit
to display zero.
Screw the control valve firmly onto the cylinder and connect the gas
supply as shown below:
Serial
port
Calibration
Switch